DOI: 10.1056/NEJM199408253310802

Background Coronary-stent placement is a new technique in which a balloon-expandable, stainless-steel, slotted tube is implanted at the site of a coronary stenosis. The purpose of this study was to compare the effects of stent placement and standard balloon angioplasty on angiographically detected restenosis and clinical outcomes. Methods We randomly assigned 410 patients with symptomatic coronary disease to elective placement of a Palmaz-Schatz stent or to standard balloon angioplasty. Coronary angiography was performed at base line, immediately after the procedure, and six months later. Results The patients who underwent stenting had a higher rate of procedural success than those who underwent standard balloon angioplasty (96.1 percent vs. 89.6 percent, P = 0.011), a larger immediate increase in the diameter of the lumen (1.72 ±0.46 vs. 1.23 ±0.48 mm, P<0.001), and a larger luminal diameter immediately after the procedure (2.49 ±0.43 vs. 1.99 ±0.47 mm, P<0.001). At six months, the patients with stented ...